ORIGINAL ARTICLE
Year : 2020  |  Volume : 10  |  Issue : 2  |  Page : 145-151

Influence of kinematics on the cyclic fatigue of ProTaper Gold and WaveOne Gold


1 Department of Endodontics, King Fahad General Hospital, Jeddah, Saudi Arabia
2 Department of Restorative Dentistry, Faculty of Dentistry, King Abdulaziz University, Jeddah, Saudi Arabia

Date of Submission24-Aug-2019
Date of Decision20-Sep-2019
Date of Acceptance01-Oct-2019
Date of Web Publication23-Apr-2020

Correspondence Address:
Dr. Nawaf S Hanbazaza
Department of Endodontics, King Fahad General Hospital, Jeddah
Saudi Arabia
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/sej.sej_132_19

Rights and Permissions
  Abstract 

Introduction: The aim of the study is to assess the difference in the dynamic cyclic fatigue resistance of ProTaper Gold (PTG) and WaveOne Gold (WOG) files tested in clockwise and counterclockwise continuous rotation against two different reciprocal motions.
Materials and Methods: A total of 200 files were tested at 300 rpm. These files were divided into four groups according to direction and type of motion: (A) clockwise continuous rotation, (B) counterclockwise continuous rotation, (C) reciprocal motion with 150° counterclockwise and then 30° clockwise motion, and (D) reciprocal motion with 150° clockwise and then 30° counterclockwise motion. The files were operated in a stainless steel artificial canal with a 60° angle of curvature and a 5-mm radius. A continuous, axial oscillating motion was applied at one cycle/s. The number of cycles to failure (NCF) was recorded. The data were analyzed using two-way analysis of variance.
Results: No significant difference in NCF was found between Groups A and B or between Groups C and D in all rotary systems used (P > 0.05). However, reciprocal motions exhibited higher cyclic fatigue resistance than continuous rotations (P < 0.05). Moreover, PTG showed higher cyclic fatigue resistance than WOG (P < 0.05).
Conclusions: Reciprocal motions have increased cyclic fatigue resistance, but the direction of the motion had no effect on the fatigue life of the tested instruments. PTG files had higher resistance to cyclic fatigue than WOG files.

Keywords: Cyclic fatigue, kinematics, ProTaper Gold, WaveOne Gold


How to cite this article:
Hanbazaza NS, Abuhaimed TS. Influence of kinematics on the cyclic fatigue of ProTaper Gold and WaveOne Gold. Saudi Endod J 2020;10:145-51

How to cite this URL:
Hanbazaza NS, Abuhaimed TS. Influence of kinematics on the cyclic fatigue of ProTaper Gold and WaveOne Gold. Saudi Endod J [serial online] 2020 [cited 2020 Aug 10];10:145-51. Available from: http://www.saudiendodj.com/text.asp?2020/10/2/145/283136


  Introduction Top


Stainless steel instruments have long been used for root canal treatment. Craig et al.[1],[2] demonstrated that stainless steel files were more resistant to cyclic fatigue and had similar or improved resistance to torsional failure than carbon steel instruments. These characteristics and the resistance of stainless steel to corrosion during sterilization procedures led to the use of stainless steel files as the primary endodontic file.[3] However, these files had a tendency to introduce procedural errors that altered the natural canal anatomy. Instrument separation was an additional hazard.[4] One reason for the difficulty in maintaining the original canal shape with stainless steel files was because these files had limited flexibility, particularly when exceeding an ISO size of #35.[1]

In the 1960s, Buehler and Wang[5] developed a Nitinol wire while working at the Naval Ordnance Laboratory in White Oak, Maryland. They derived the name Nitinol from nickel and titanium and NOL for the Naval Ordnance Laboratory. This wire is now more commonly referred to as nickel titanium (NiTi). In 1988, Walia et al.[6] milled Nitinol orthodontic wires as prototype endodontic files and compared their bending and torsional properties to stainless steel files. In their research, they found that the NiTi files were two to three times more flexible in bending and torsion and more resistant to torsional failure.

Rotary instruments made of NiTi alloys were commonly used in endodontic treatments to shape root canals with challenging anatomies. Cyclic fatigue was the most common cause of fractured NiTi instruments.[7]

Fatigue fractures are caused by a repeated cycle of tension and compression at the maximum curved part of the instrument, as this initiates a crack on the surface and then the crack increases.[7] Increasing resistance to this cyclic fatigue has been the focus for advancement of rotary NiTi instrument technology. Improvements in the manufacturing process are among the strategies proposed to increase the mechanical properties of these NiTi files.

The motion type and the direction of the movements affect the instruments' fatigue resistance. Most rotary instruments are used in a clockwise continuous rotation, but that motion increases the stresses on the file, particularly in curved and narrow canals.[7] Reciprocal movements were introduced to overcome this shortcoming.[8]

As proposed by Yared,[8] a reciprocal motion is defined as the movement of the file in the clockwise direction followed by a counterclockwise releasing movement. Recent publications have claimed that reciprocation increases the resistance of NiTi files to cyclic fatigue and decreases the incidence of instrument separation during root canal preparation.[9] Limited information is available about the influence of clockwise continuous rotation, counterclockwise continuous rotation, and reciprocation on the cyclic fatigue of rotary systems.

The aim of the study was to evaluate the resistance of ProTaper Gold (PTG) and WaveOne Gold (WOG) NiTi files to dynamic cyclic fatigue and to assess the lengths of broken segments in clockwise continuous rotation, counterclockwise continuous rotation, clockwise reciprocal motion, and counterclockwise reciprocal motion.


  Materials and Methods Top


A laboratory experiment for testing the dynamic cyclic fatigue of two different rotary systems was conducted by one operator at a dental biomaterials research laboratory at King Abdulaziz University, Jeddah, Saudi Arabia (Ethical approval number: 016:01–17).

A total of 200 files were tested. The files were divided into four groups according to the directions and the type of motion (n = 50):

  • Group A: Clockwise continuous motion
  • Group B: Counterclockwise continuous motion
  • Group C: Reciprocal motion with 150° clockwise and 30° counterclockwise m otion
  • Group D: Reciprocal motion with 150° counterclockwise and 30° clockwise motion.


Each group was subdivided into two subgroups according to the rotary system used (n = 25):

  • 1: PTG, F1, tip size # 20 and 7% taper (Dentsply Maillefer, Ballaigues, Switzerland)li>
  • 2: WOG, tip size # 20 and 7% taper (Dentsply Maillefer, Ballaigues, Switzerland).


Each file was examined for defects before the test with a dental operating microscope (PICO, Carl Ziess, Jena, Germany).

A special setup was assembled for the current experiment using a universal testing machine (MultiTest 2.5-i, Mecmesin, Slinfold, UK) [Figure 1]. A contra-angle handpiece with 1:16 reduction (Sirona, 64625 Bensheim, Germany) and electric endodontic motor (Setelec I-Endo Dual, Acteon, France) was used. The handpiece of the endodontic motor was attached to the moving arm of the universal testing machine. At the lower end of the machine, a stainless steel artificial canal was attached to a chuck vice, which was fixed to a two-dimensional horizontal micropositional stage to precisely align the artificial canal with the descending file [Figure 1].
Figure 1: Universal testing machine (1: The moving arm of the testing machine, 2: A contra-angle handpiece with 1:16 reduction, 3: Stainless steel artificial canal, 4: Chuck vice, and 5 and 6: A two-dimensional horizontal micropositional stage to precisely align the artificial canal with the file)

Click here to view


The artificial canal was made of a laser milled stainless steel block to standardize the testing method. The angle of the artificial canal was 60° with a radius of 5 mm according to Pruett et al.'s[10] method [Figure 2]. The artificial canal was designed to accommodate the tested rotary files with a tip size of 20 and a 7% taper; the depth of each artificial canal was laser milled to the maximum diameter of the instrument + 0.2 mm, allowing the instrument to rotate freely inside the canal [Figure 2].
Figure 2: Laser-milled stainless steel block

Click here to view


A synthetic oil (WD-40 Company, Milton Keynes, England) was used for lubrication as recommended by Nguyen et al.[11] A drop of oil was used for each rotary file to minimize the friction between the rotating files and the stainless-steel canal wall and prevent temperature elevation. Moreover, the artificial canal was covered with glass to prevent the file and the broken segment from slipping out and for better observation of the rotating file [Figure 3].
Figure 3: Laser-milled stainless steel block covered by glass

Click here to view


All files were 25 mm in length and tested at 300 rpm, according to the manufacturer's recommended speed range.

Twenty-five files in each group were rotated freely in the stainless steel artificial canal. After each file was positioned, a video recording in a high-resolution camera (D3200; Nikon, Tokyo, Japan) was initiated. The time to fracture in seconds was calculated from the videotape and started when the file began to rotate until observing the first sign of fracture. Moreover, instrument fracture was visually detected and timed using a digital stopwatch accurate to 0.01 s.

A continuous axial oscillating motion was applied at one (hertz) cycle/s to simulate a clinical pecking motion. The file was cycled 3 mm above and 3 mm below the starting point, resulting in a net oscillation movement of 6 mm. A special software (Emperor, Mecmesin, Slinfold, UK) was used to control the axial motion.

The number of cycles to failure (NCF) for each instrument was calculated by converting the time (in seconds) required to fracture the instrument into decimal units (in minutes) multiplied by speed.

After removing the glass plate that covered the milled stainless steel block, a Vernier caliper (model 40185; Sears, Roebuck and Co., Chicago, IL, USA) was used to measure the length of the separated instrument's segments.

Statistical analysis

The NCF was analyzed statistically using two-way analysis of variance (ANOVA) followed by post hoc pairwise test (SPSS for Mac 11.0; SPSS, Chicago, IL, USA) at a 95% confidence level.

The lengths of fractured segments were analyzed statistically using one-way ANOVA and Mann–Whitney U-tests (SPSS for Mac 11.0; SPSS, Chicago, IL, USA) at a 95% confidence level.


  Results Top


The mean, standard deviation, and statistical analysis of tested groups are presented in [Table 1], [Table 2], [Table 3].
Table 1: The mean±standard deviation of the number of cycles to failure of the tested groups

Click here to view
Table 2: Pairwise comparisons (post hoc test)


Click here to view
Table 3: Pairwise comparisons (post hoc test)

Click here to view


PTG files exhibited a statistically significant higher cyclic fatigue resistance when compared to WOG (P < 0.05). Moreover, the fatigue life of Groups C and D (Reciprocal motion) were significantly higher than Groups A and B (continuous rotation motion) (P < 0.05). The direction of the motion (clockwise versus counterclockwise directions) did not influence the cyclic fatigue resistance in all groups tested (P > 0.05).

The average length of PTG fractured segments for all samples was 4.1 mm and 2.6 mm for WOG segments.

The values and a statistical analysis of the fracture fragments length are presented in [Table 4]. There was a statistical difference between PTG and WOG files (P < 0.01).
Table 4: The mean±standard deviation of fractured instrument fragment length (FI) of the groups tested

Click here to view



  Discussion Top


During root canal cleaning and shaping, choosing the instrument with higher fracture resistance decreases the incidence of instrument separation; therefore, many studies focus on the cyclic fatigue resistance of NiTi files. Manufacturers aim to improve cyclic fatigue resistance by modifying the metallurgy, kinematics, and design of NiTi files.[12],[13] Therefore, it is crucial for clinicians to understand the different physical properties, kinematics, and designs of recent rotary systems to achieve the maximum benefits of current developments.

This study evaluated the difference in dynamic cyclic fatigue resistance of PTG and WOG systems tested in different continuous rotation and reciprocal motions. To standardize cyclic fatigue testing and eliminate other contributing factors apart from the cross-sectional design, instruments with a size of 20 and a 7% taper from PTG and WOG were used in this study.

Major drawbacks of laboratory studies testing the cyclic fatigue resistance of NiTi rotary files were the contributing variables that cannot be completely eliminated, which make it difficult to measure the effect of a single variable on its resistance to cyclic fatigue.[14] The clinical relevance of these laboratory study results was difficult to assess because the condition differs from clinical intracanal instrumentation in which the file separation might occur due to combined factors (cyclic fatigue and torsional stress) that act simultaneously.[15] Yao et al.[16] stated that “Although the use of extracted teeth simulates clinical situations, they were not anatomically standardized.” Therefore, extracted teeth were not optimal for testing the cyclic fatigue resistance for NiTi instruments. In this study, an artificial canal was used to rule out all the other variables other than the cyclic fatigue variable.

In the literature, the cyclic fatigue resistance of rotary NiTi files has been examined in several static and dynamic models.[17],[18],[19],[20],[21],[22],[23],[24],[25],[26] These include rotating the NiTi files against an inclined plane by means of a grooved block or a curved metal tube.[26],[27] Cyclic fatigue test devices could be used in either static or dynamic modes. In static mode, the instrument rotates in the fixed length in the artificial canal with no axial movement.[28] In dynamic mode, the instrument rotates in the artificial canal with axial movement. Dynamic models were reported to simulate clinical situations more realistically than static models.[19],[29]

In the present study, PTG had a significantly higher resistance to cyclic fatigue than WOG in all types of motion tested. This finding could be due to the differences in the cross-sectional design. WOG files have a parallelogram-shaped design with cutting edges and alternate one-point contact, whereas PTG files have convex triangular cross sections.[17],[30] Both PTG and WOG files have constant tapers between D1 and D3, which decreases progressively from D4 to D14.[31],[32] Moreover, both systems were manufactured by gold heat treatment.[17],[30] A cross-sectional design has been proven to modify the stress distribution produced by a rotary instrument under twisting or tension.[14],[33] Cheung et al.[14] conducted a study using a finite elemental analysis and proved that a triangular cross-sectional design showed a higher cyclic fatigue resistance than a square cross-sectional design. They attributed the higher cyclic fatigue resistance of triangular cross-sectional instruments to less metal mass of the files with a triangular cross-section compared to a square cross-section of a similar diameter.[34],[35] Hence, in this study, the higher resistance to cyclic fatigue of PTG files might be due to the smaller metal mass of the triangular cross compared with more metal in the parallelogram-shaped cross section in the WOG system.

In the present study, the reciprocal motions tested had significantly higher resistance to cyclic fatigue than continuous rotations in both clockwise and counterclockwise directions.

De-Deus et al.[9] evaluated the cyclic fatigue resistance of ProTaper F2 instruments during continuous rotation and reciprocal motion. They reported that reciprocal motion had significantly higher resistance to cyclic fatigue than continuous rotation motion. The fatigue life is measured by how many times that the initiated crack opens and closes. During one continuous rotation cycle, the crack opens and closes once. According to Yared,[8] during one reciprocating movement, the file moved 144° clockwise then moved back 72° counterclockwise, which means that five reciprocal movements will be required to complete one full cycle. In this study, during each reciprocating movement, the files turned 150° then turned back 30°, which means that three reciprocal movements will be required to complete one full cycle. Hence, it will require at least three reciprocal movements to open and close the initiated crack, which is the reason for a higher resistance to cyclic fatigue in reciprocal motions than in continuous rotation motions.

In this study, the time required for different motions to complete one full rotation with a speed of 300 rotations/min was recorded. For continuous rotation, a 0.216 s was recorded to complete one full rotation. However, for reciprocal motion, a 0.288 s is required to complete one 360° full rotation. In other words, reciprocal motion moves at a lower speed to complete a full rotation compared to continuous rotation. Therefore, reciprocal motion had higher resistance to cyclic fatigue because it required a longer time for the initiated crack to go through one cycle of compression and tension.

However, the present study showed that there was no significant difference in the cyclic fatigue resistance between a clockwise direction and a counterclockwise direction in both reciprocal and continuous rotation motions, and this might be due to identical stress distribution points during bending moments in clockwise and counterclockwise directions.

This study showed that there was no significant difference in the mean length of the fractured segments between each type of motion. Gündoǧar and Özyürek[21] tested the cyclic fatigue of WOG and Reciproc Blue rotary systems. They reported that there was no significant difference between the mean length of the fractured fragments of the rotary systems used. They attributed that to the precise and reproducible position of each file inside the canal during testing. In this study, the mean length of the fractured segments was not significantly different between all types of motion tested in each rotary file system. This might be because each file tested was accurately positioned inside the canal.

On the other hand, the mean length of fractured segments of PTG files tested in this study was 4.1 mm, which was at the center of the curvature. However, the mean length of fractured segments of WOG files tested in this study was 2.4 mm, which was below the center of the curvature. There was a significant difference in the mean lengths of the fractured segments between PTG and WOG. Kaval et al.[36] compared the cyclic fatigue resistance of ProTaper Universal and PTG systems. They reported that the mean length of PTG was 5.9 mm, which was significantly higher than the mean length (4.9 mm) of ProTaper universal. Previous studies attributed the difference in fragment lengths among rotary systems to the physical properties of the alloy and the cross-sectional design of the files, which could modify the distribution of the maximum stress points.[36],[37],[38],[39]

During cyclic fatigue testing, each point at the curved part of the file was bent and went through compression and tension cycles. During bending moments, the crack was initiated at the point where the maximum stress was located and propagated until fracture occurs. The length of fractured segment gave us an indication of the maximum stress point.

In the present study, the mean length of fractured segments of the WOG files was observed to be shorter than the PTG files, which might be due to different locations of the maximum stress point during compression, and the tension cycles due to the different cross-sectional designs of the PTG and WOG files.

In dynamic mode, the time to fracture scores increased because the stress was distributed along the shaft during back and forth axial movements when compared with static test modes in which the stress was concentrated in a specific area.[23],[28],[29] In this study, the dynamic model for testing cyclic fatigue was used to represent the clinical situation as much as possible.


  Conclusions Top


PTG files had higher dynamic cyclic fatigue resistance than WOG files due to the differences in their cross-sectional designs. Reciprocal motion increased cyclic fatigue resistance, but the direction of motion had no effect on the cyclic fatigue resistance of the tested instruments. Kinematics had no influence on the length of separated segments. However, PTG had longer separated segments than WOG due to the differences in their geometrical designs.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Craig RG, McIlwain ED, Peyton FA. Bending and torsion properites of endodontic instruments. Oral Surg Oral Med Oral Pathol 1968;25:239-54.  Back to cited text no. 1
    
2.
Craig RG, Peyton FA. Physical properties of stainless steel endodontic files and reamers. Oral Surg Oral Med Oral Pathol Oral Radiol 1963;16:206-17.  Back to cited text no. 2
    
3.
Younis O. The effects of sterilization techniques on the properties of intracanal instruments. Oral Surg Oral Med Oral Pathol 1977;43:130-4.  Back to cited text no. 3
    
4.
Weine FS, Kelly RF, Lio PJ. The effect of preparation procedures on original canal shape and on apical foramen shape. J Endod 1975;1:255-62.  Back to cited text no. 4
    
5.
Buehler WJ, Wang FE. A summary of recent research on the nitinol alloys and their potential application in ocean engineering. Ocean Eng 1968;1:105-20.  Back to cited text no. 5
    
6.
Walia HM, Brantley WA, Gerstein H. An initial investigation of the bending and torsional properties of nitinol root canal files. J Endod 1988;14:346-51.  Back to cited text no. 6
    
7.
Parashos P, Gordon I, Messer HH. Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use. J Endod 2004;30:722-5.  Back to cited text no. 7
    
8.
Yared G. Canal preparation using only one Ni-Ti rotary instrument: Preliminary observations. Int Endod J 2008;41:339-44.  Back to cited text no. 8
    
9.
De-Deus G, Moreira EJ, Lopes HP, Elias CN. Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J 2010;43:1063-8.  Back to cited text no. 9
    
10.
Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 1997;23:77-85.  Back to cited text no. 10
    
11.
Nguyen HH, Fong H, Paranjpe A, Flake NM, Johnson JD, Peters OA. Evaluation of the resistance to cyclic fatigue among ProTaper next, ProTaper universal, and vortex blue rotary instruments. J Endod 2014;40:1190-3.  Back to cited text no. 11
    
12.
Xu X, Eng M, Zheng Y, Eng D. Comparative study of torsional and bending properties for six models of nickel-titanium root canal instruments with different cross-sections. J Endod 2006;32:372-5.  Back to cited text no. 12
    
13.
Park SY, Cheung GS, Yum J, Hur B, Park JK, Kim HC. Dynamic torsional resistance of nickel-titanium rotary instruments. J Endod 2010;36:1200-4.  Back to cited text no. 13
    
14.
Cheung GS, Zhang EW, Zheng YF. A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments. Int Endod J 2011;44:357-61.  Back to cited text no. 14
    
15.
Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod 2009;35:1469-76.  Back to cited text no. 15
    
16.
Yao JH, Schwartz SA, Beeson TJ. Cyclic fatigue of three types of rotary nickel-titanium files in a dynamic model. J Endod 2006;32:55-7.  Back to cited text no. 16
    
17.
Elnaghy AM, Elsaka SE. Effect of sodium hypochlorite and saline on cyclic fatigue resistance of WaveOne gold and reciproc reciprocating instruments. Int Endod J 2017;50:991-8.  Back to cited text no. 17
    
18.
Özyürek T. Cyclic fatigue resistance of reciproc, WaveOne, and WaveOne gold nickel-titanium instruments. J Endod 2016;42:1536-9.  Back to cited text no. 18
    
19.
Topçuoǧlu HS, Düzgün S, Aktı A, Topçuoǧlu G. Laboratory comparison of cyclic fatigue resistance of WaveOne gold, reciproc and WaveOne files in canals with a double curvature. Int Endod J 2017;50:713-7.  Back to cited text no. 19
    
20.
Adıgüzel M, Capar ID. Comparison of cyclic fatigue resistance of waveOne and waveOne gold small, primary, and large instruments. J Endod 2017;43:623-7.  Back to cited text no. 20
    
21.
Gündoǧar M, Özyürek T. Cyclic Fatigue Resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc Blue Nickel-titanium Instruments. J Endod 2017;43:1192-6.  Back to cited text no. 21
    
22.
Keskin C, Inan U, Demiral M, Keleş A. Cyclic fatigue resistance of reciproc blue, reciproc, and waveOne gold reciprocating instruments. J Endod 2017;43:1360-3.  Back to cited text no. 22
    
23.
Özyürek T, Yılmaz K, Uslu G. Effect of adaptive motion on cyclic fatigue resistance of a nickel titanium instrument designed for retreatment. Restor Dent Endod 2017;42:34-8.  Back to cited text no. 23
    
24.
Keskin N, Özyurek T, Uslu G, Inan U. Cyclic fatigue resistance of new and used ProTaper universal and ProTaper next nikcel-titanium rotary instruments. Saudi Endod J 2017;7:151-5.  Back to cited text no. 24
    
25.
Uslu G, Yilmaz K, Özyurek T. Influence of different movement kinematics on cyclic fatigue resistance of nickel-titanium instruments designed for retreatment. Saudi Endod J 2018;8:82-6.  Back to cited text no. 25
  [Full text]  
26.
Abuhaimed T. Effect of environment on cyclic fatigue of ProTaper next rotary files. Saudi Endod J 2018;8:117-22.  Back to cited text no. 26
  [Full text]  
27.
Cheung GS, Darvell BW. Low-cycle fatigue of NiTi rotary instruments of various cross-sectional shapes. Int Endod J 2007;40:626-32.  Back to cited text no. 27
    
28.
Lopes HP, Elias CN, Vieira MV, Siqueira JF Jr. Mangelli M, Lopes WS, et al. Fatigue life of reciproc and mtwo instruments subjected to static and dynamic tests. J Endod 2013;39:693-6.  Back to cited text no. 28
    
29.
Pedullà E, Lizio A, Scibilia M, Grande NM, Plotino G, Boninelli S, et al. Cyclic fatigue resistance of two nickel-titanium rotary instruments in interrupted rotation. Int Endod J 2017;50:194-201.  Back to cited text no. 29
    
30.
Elnaghy AM, Elsaka SE. Mechanical properties of ProTaper gold nickel-titanium rotary instruments. Int Endod J 2016;49:1073-8.  Back to cited text no. 30
    
31.
DentSply Sirona WaveOne® Gold Endodontics: Brochure. Available from: https://www.dentsplysirona.com/en-ae/explore/endodontics/waveone-gold-reciprocating-files.html. [Last accessed on 2018 May 20].  Back to cited text no. 31
    
32.
DentSply Sirona. ProTaperGold™ Gold Endodontics: Brochure. Available from: https://www.dentsplysirona.com/en-ae/explore/endodontics/protaper-gold-files.html. [Last accessed on 2018 May 20].  Back to cited text no. 32
    
33.
Zhang EW, Cheung GS, Zheng YF. Influence of cross-sectional design and dimension on mechanical behavior of nickel-titanium instruments under torsion and bending: A numerical analysis. J Endod 2010;36:1394-8.  Back to cited text no. 33
    
34.
Uygun AD, Kol E, Topcu MK, Seckin F, Ersoy I, Tanriver M. Variations in cyclic fatigue resistance among ProTaper gold, ProTaper next and ProTaper universal instruments at different levels. Int Endod J 2016;49:494-9.  Back to cited text no. 34
    
35.
Capar ID, Kaval ME, Ertas H, Sen BH. Comparison of the cyclic fatigue resistance of 5 different rotary pathfinding instruments made of conventional nickel-titanium wire, M-wire, and controlled memory wire. J Endod 2015;41:535-8.  Back to cited text no. 35
    
36.
Kaval ME, Capar ID, Ertas H. Evaluation of the cyclic fatigue and torsional resistance of novel nickel-titanium rotary files with various alloy properties. J Endod 2016;42:1840-3.  Back to cited text no. 36
    
37.
Capar ID, Ertas H, Arslan H. Comparison of cyclic fatigue resistance of nickel-titanium coronal flaring instruments. J Endod 2014;40:1182-5.  Back to cited text no. 37
    
38.
Hussne RP, Braga LC, Berbert FL, Buono VT, Bahia MG. Flexibility and torsional resistance of three nickel-titanium retreatment instrument systems. Int Endod J 2011;44:731-8.  Back to cited text no. 38
    
39.
Shen Y, Hieawy A, Huang X, Wang ZJ, Maezono H, Haapasalo M. Fatigue resistance of a 3-dimensional conforming nickel-titanium rotary instrument in double curvatures. J Endod 2016;42:961-4.  Back to cited text no. 39
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Materials and Me...
Results
Discussion
Conclusions
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed219    
    Printed19    
    Emailed0    
    PDF Downloaded67    
    Comments [Add]    

Recommend this journal