Cyclic fatigue resistance of three heat-treated nickel-titanium instruments at simulated body temperature


1 Endodontic Department, Faculty of Dentistry, Ain Shams University, Cairo, Egypt
2 Endodontic Department, Faculty of Dentistry, Ain Shams University; Department of Endodontics, The British University, Cairo, Egypt
3 Endodontics Unit, Department of Oral and Maxillo-Facial Sciences, “Sapienza” University of Rome, Italy

Date of Submission08-Aug-2019
Date of Decision18-Sep-2019
Date of Acceptance25-Sep-2019
Date of Web Publication23-Apr-2020

Correspondence Address:
Dr. Tarek M Elsewify
Department of Endodontics, Faculty of Dentistry, Ain Shams University, Cairo
Egypt
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/sej.sej_122_19

Rights and Permissions
  Abstract 

Introduction: The aim of this study was to evaluate the cyclic fatigue resistance of three heat-treated nickel–titanium (NiTi) systems at simulated body temperature.
Materials and Methods: Twelve instruments of similar apical diameter (#25) from three engine-driven NiTi instruments; WaveOne Gold (WOG), Hyflex EDM (HEDM), and M Pro system were tested for cyclic fatigue resistance at 37°C in simulated root canals with 60° angle of curvature and 5 mm curvature radius. All instruments were operated until fracture occurred. Mean and standard deviation of the time to fracture and fragment length were calculated, and statistical significance was set at 5%. The morphological characteristics of the fractured instruments were observed through scanning electron microscopy, and their chemical composition was determined using energy dispersive X-ray analysis (EDXA).
Results: HEDM showed the highest resistance to cyclic fatigue at 37°C followed by WOG and M Pro instruments with a statistically significant difference. Scanning electron micrographs confirmed a predominantly ductile mode of fracture for all instruments. EDXA showed that WOG was composed of Ni (45.1 wt%) and Ti (37.0 wt%), HEDM was composed of Ni (52.9 wt%) and Ti (42.4 wt%), and M Pro was composed of Ni (49.7 wt%) and Ti (40.4 wt%).
Conclusions: M Pro instrument showed the least cyclic fatigue resistance at 37°C compared to HEDM and WOG.

Keywords: Cyclic fatigue, Hyflex EDM, M Pro, nickel-titanium, WaveOne Gold


How to cite this article:
Elsewify TM, Saber SM, Plotino G. Cyclic fatigue resistance of three heat-treated nickel-titanium instruments at simulated body temperature. Saudi Endod J 2020;10:131-6

How to cite this URL:
Elsewify TM, Saber SM, Plotino G. Cyclic fatigue resistance of three heat-treated nickel-titanium instruments at simulated body temperature. Saudi Endod J [serial online] 2020 [cited 2020 Aug 10];10:131-6. Available from: http://www.saudiendodj.com/text.asp?2020/10/2/131/283134


  Introduction Top


Nickel–titanium (NiTi) mechanical instruments are able to shape root canals faster, with better-centered preparations and less procedural errors than stainless-steel ones.[1] Yet, unexpected instrument fracture still does occur.[2] Two modes of fracture were identified by Sattapan et al.:[3] torsional failure and cyclic fatigue. Torsional failure occurs upon reaching the ultimate shear strength. Cyclic fatigue is attributed to metal fatigue when it rotates freely in a curved canal at the point of maximum flexure.[4],[5] Factors contributing to the cyclic fatigue of mechanical NiTi instruments rotating in a curvature include root canal anatomy (radius and degree of the curvature),[6] cleaning and shaping techniques,[7] debris accumulation[8] and sterilization procedures,[9] size,[3] taper,[10] cross-section,[11] and instrument design,[12] as well as the manufacturing technique.[13] Unfortunately, cyclic fatigue often occurs without any visible sign of plastic deformation.[14]

Introduction of new alloys in addition to the newly developed manufacturing processes aimed to improve the cyclic fatigue resistance of NiTi instruments.[14] Reciprocation motion was also shown to improve the cyclic fatigue resistance.[15],[16],[17]

WaveOne Gold (WOG) instruments are the updated version of WaveOne files (Dentsply Tulsa Dental; Tulsa, OK, USA). They are still used in reciprocating motion although with different dimensions, geometry, and cross-section. They possess a parallelogram cross-section with two cutting edges and an off-center design.[18] Gold heat treatment is done through heating followed by slowly cooling of the produced file, unlike M-Wire technology which is done before production.[19]

Hyflex EDM (HEDM) (Coltene/Whaledent; Altstatten, Switzerland) is a single-file system used in continuous rotary motion. Controlled memory (CM) alloy is used for their manufacturing using electric discharge machining technology. The mechanical properties of the HEDM instruments were shown to be significantly improved because of CM wire and method of manufacturing.[18]

M Pro rotary (IMD; ShangHai, China) NiTi instruments are made of X wire, a CM alloy with a special heat treatment to improve flexibility and strength.[20] They showed a convex triangular cross-section with three cutting edges and large core and a progressively increasing pitch along with the blades.[20] No study examining the cyclic fatigue resistance of the M Pro rotary NiTi instrument could be found in the literature. Thus, the aim of the current study was to compare the cyclic fatigue resistance of WOG, HEDM, and M Pro NiTi instruments at simulated body temperature.


  Materials and Methods Top


Sample selection

Three types of engine-driven NiTi instruments were tested: WOG (Dentsply-Sirona Endodontics, Baillagues, Switzerland), Hyflex EDM (Coltene/Whaledent, Altstatten, Switzerland), and M Pro (IMD, ShangHai, China).

Statistical sample size calculation was performed using the following formula:

N = (Zα)2 S2/d2, where Z value is 1.64 for 5% level of significance, s is the standard deviation obtained from a previous study,[18] and d is the accuracy of estimate. For each system, 12 instruments of similar tip dimension (# 25) were tested for their cyclic fatigue resistance using a mechanical device used earlier in a previous study.[21]

Cyclic fatigue test

Instruments were operated according to manufacturer's instructions in a simulated stainless steel canal using a 6:1 reduction handpiece (Reciproc silver, VDW, Germany). The simulated root canal was 18 mm long with a 60° angle of curvature and a 5-mm radius of curvature according to Pruett's method.[4] The curved segment of the canal was 6 mm and the center of curvature was 6 mm away from the instrument tip. After accurate instrument positioning to its full length, it was rotated freely and synchronized with timing by a digital stopwatch (Timex, Middlebury, CT, USA) to the thousandth of a second. The cyclic fatigue apparatus was inserted in a thermocycler water bath at 37°C, and testing of all instruments was performed.

“WaveOne All” program was used for the WOG primary files (tip size 25/0.07 taper in the apical 3 mm and variable taper more coronally). HEDM files (tip size 25/variable taper) were rotated at 400 rpm and torque of 3 Ncm, while M Pro files (tip size 25/0.06 taper) were rotated at a constant speed of 400 rpm and torque of 1.5 Ncm. All tested instruments were rotated until fracture occurred. Timing was initiated as soon as the motion started after accurate positioning of the tested instrument into the canal. Time was calculated till fracture as detected visually and audibly. A ×2.5 magnifying binocular loupes with LED light (Heine; Herrsching, Germany) was used to determine the failure time. Testing of all instruments was performed by one operator. For each instrument, the time to fracture (TtF) in seconds was recorded. The length of the separated fragment was also measured for each instrument with a digital caliper.

Statistical analysis

Mean and standard deviation (SD) of data obtained were calculated for each group and statistically analyzed using one-way analysis of variance and post hoc Tukey test with SPSS version 18.0 (SPSS Inc, Chicago, IL, USA). Statistical significance was set at 5%.

Scanning electron microscopic evaluation

Random selection of three instruments from each group was made. These instruments were ultrasonically cleaned in absolute alcohol for approximately 120 s and then mounted for fractographic examination by scanning electron microscopy (Philips SEM 515, Eindhoven, Netherlands) at ×200 and ×3000 magnification to evaluate the type of fracture with energy dispersive X-ray analysis (EDXA) (Philips SEM 515, Eindhoven, The Netherlands) made to identify their metallurgic composition. Analyses were done at 25 kV at room temperature.


  Results Top


Mean TtF ± SD values after cyclic fatigue testing of WOG, HEDM, and M Pro NiTi instruments are displayed in [Table 1]. One-way ANOVA revealed a significant difference among the three tested instruments (P < 0.05). HEDM showed the highest resistance to cyclic fatigue at simulated body temperature followed by the WOG and the M Pro instruments.
Table 1: Mean time to fracture in seconds (standard deviation) and fragment length in mm after cyclic fatigue testing of WaveOne Gold, Hyflex EDM, and M Pro rotary files in 60° curved simulated canals

Click here to view


SEM of the fractured surface of the tested instruments confirmed a ductile fracture, evident by the numerous dimples on the fractured surface. These dimples resulted from the formation of microvoids in the center of the metal shaft (overload zone), as it underwent cyclic fatigue. At high magnification (×3000), the dimpled surfaces revealed a characteristic population of microvoids for each instrument type. For HEDM and M Pro instruments, the dimples were fewer and relatively smaller. For WOG instruments, the dimples were numerous and relatively larger [Figure 1].
Figure 1: Scanning electron micrographs of the fractured surfaces obtained from the instruments after cyclic fatigue testing. At low magnification (×200; a, c, and e), centrally located overload zones with evidence of multiple crack initiation sites (black arrows) around the perimeter of the instrument are shown. At high magnification (×3000; b, d, and f), overload zones revealed evidence of dimpled rupture that occurred due to the coalescence of the microvoids in the overload zone, resulting in the ultimate ductile fracture of the instruments. A large number of microvoids (red arrows) could be seen within the core of the instrument

Click here to view


EDXA showed that WOG instruments were composed of nickel (45.1 wt%), titanium (37.0 wt%), and carbon (15.5 wt%). HEDM instruments were composed of nickel (52.9 wt%), titanium (42.4 wt%), and nitrogen (4.5 wt%). M Pro instruments were composed of nickel (49.7 wt%), titanium (40.4 wt%), oxygen (8.5wt%), and traces of silicon (1.2 wt%) as shown in [Figure 2].
Figure 2: Graphic representation of the energy dispersive X-ray analysis results showing the elemental composition of WaveOne Gold (a), Hyflex EDM (b), and M Pro (c) files. KV: 25.00 Tilt: 0.00 Take-off: 35.00 Ampt: 102.4 Detector type: Sutw-Sapphire resolution: 129.55

Click here to view



  Discussion Top


Attempts to improve the cyclic fatigue resistance of NiTi instruments were made by changing the metallurgy, design, and kinematics of the instruments.[15] The aim of this study was to evaluate the cyclic fatigue resistance of M Pro NiTi instruments. Pruett's method was chosen for this study to define root canal geometry.

The current study was designed to overcome limitations of otherin vitro studies through using specifically designed artificial canals for each instrument, thus giving it a precise trajectory during testing.[21],[22] Yet, we do not have a standardized technique for cyclic fatigue resistance evaluation of rotary NiTi instruments. The ideal technique would use natural curved root canals. However, the tooth will be used only once and standardization of the experimental conditions will be impossible.[23]

A dynamic model for cyclic fatigue testing represents the clinical situation more than a static one; yet, it is more liable to procedural errors and is more difficult to keep the instruments in a precise trajectory.[24]

It has been clearly demonstrated that temperature has a significant effect on cyclic fatigue resistance of all mechanical NiTi endodontic instruments.[25],[26],[27] Increase in temperature affects the properties of NiTi alloys.[28] Rotary instruments work at an intracanal temperature of 31°C–35°C[29] rather than room temperature. It was clearly demonstrated that the increase in temperature from room temperature to body temperature affects negatively the cyclic fatigue resistance of rotary NiTi instruments.[24],[26],[30] This could be explained on the basis of the metallurgical properties of the endodontic instruments. Predominantly, martensitic instruments at room temperature are transformed to a more austenitic phase upon heating, thus decreasing their cyclic fatigue resistance. This is related to the thermomechanical history of the NiTi alloy tested.[30] For this reason, the present study was conducted at body temperature to better simulate the clinical conditions.

The fractured segments' lengths in all groups did not show any significant difference. This indicates correct instruments positioning inside the artificial canal.[18]

M Pro showed the least cyclic fatigue resistance at simulated body temperature. This cannot be directly compared to earlier studies because none could be found in the literature.

Instruments' cross-sectional design, taper, pitch, alloy, manufacturing method, heat treatment, kinematics, and usage speed influence its cyclic fatigue resistance.[11],[15],[16],[17] The convex triangular cross-section of M Pro possesses more metal mass, and consequently, more stresses concentration than the quadratic cross-section of HEDM and the parallelogram cross-section of WOG. Another contributing factor is the constant taper of M Pro of 6%, compared to the regressing taper of HEDM and WOG, which allows for more instrument-dentin contact, more stress, and cyclic fatigue failure.

WOG showed a superior cyclic fatigue resistance to M Pro which could be easily explained on the basis of kinematics as reciprocation motion increases the cyclic fatigue resistance compared to rotation motion.[15],[31],[32] Stress concentration at one point of the instrument occurs during continuous rotation motion, rather than at multiple points at each cycle during reciprocation motion. This attenuates the magnitude of stresses built up inside the instrument during operation. An instrument rotating in a curved root canal suffers from alternating compressive and tensile stresses. Upon reciprocation, lower stresses are induced because of the shorter distance traversed in each cycle. This could be added to the gold heat treatment applied in the manufacturing of WOG instruments that was shown to increase its flexibility,[18] and thus increasing its cyclic fatigue resistance.

HEDM showed a superior cyclic fatigue resistance to M Pro as well. Although both files are CM wires used in continuous rotation, they differ in the manufacturing process. HEDM is machined by electrodischarge machining which is a noncontact thermal erosion process that partially melts and evaporates the NiTi wire by high-frequency spark discharges. This technique was proved to enhance the mechanical properties greatly.[33]

Our results showed superior fatigue resistance of HEDM to WOG which come in agreement with Pedullà et al.[34] and Gündoǧar and Özyürek.[18] This clearly demonstrates that kinematics is not the only factor that influences cyclic fatigue, but rather the cross-section, alloy type, manufacturing technique, heat treatment, and usage speed influence the cyclic fatigue life of the instrument.[11],[15],[16],[17] HEDM possesses 8% taper in the apical region and 4% in the coronal region. WOG has 7% taper apically and 3% coronally.

The difference in cross-sectional design might contribute as well. As the tested instruments fractured at D5 to D6 (mean fragment length 5.7 ± 0.5 mm), HEDM changes its cross-section from quadratic to triangular which bears less metal mass and less stress concentration than the parallelogram cross-section of WOG.

Hülsmann et al.[24] has shown that cyclic fatigue testing lacks reproducibility as while comparing a particular instrument, WaveOne, for example, a maximal difference between the values published of 200.9% is found. In most of these studies, no clear explanation could be found.

EDXA showed that HEDM instruments were composed of nickel (52.9 wt%), M Pro (49.7 wt%), and WOG (45.1 wt%). The Ni content of a NiTi alloy influences the mechanical properties of endodontic instruments.[35] With the reduction of the Ni content, there is an increased tendency to obtain stable martensite (which is more flexible than austenite) at the working temperature.[36],[37] However, it seems that the Ni content was not the most important factor in the superior behavior of HEDM.

The three tested instruments possess different metallurgic properties, manufacturing techniques, heat treatments, cross-section designs, tapers, and kinematics. Therefore, no single parameter could explain our results. Yet, we can conclude that M Pro possesses the least cyclic fatigue resistance followed by WOG and HEDM. Moreover, it is worth mentioning that the clinical situation involves a complex type of stresses, and further studies are deemed necessary to investigate the microstructure and behavior of these instruments.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Bukhari KA, Almalki MA, Daghestani MH, Bogari DF, Aljifan MK, Alharbi YM, et al. Cyclic fatigue comparison of different manufactured endodontic files. Saudi Endod J 2019;9:186-91.  Back to cited text no. 1
  [Full text]  
2.
Parashos P, Messer HH. Rotary niTi instrument fracture and its consequences. J Endod 2006;32:1031-43.  Back to cited text no. 2
    
3.
Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod 2000;26:161-5.  Back to cited text no. 3
    
4.
Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 1997;23:77-85.  Back to cited text no. 4
    
5.
Gambarini G, Seracchiani M, Piasecki L, Valenti Obino F, Galli M, Di Nardo D, et al. Measurement of torque generated during intracanal instrumentation in vivo. Int Endod J 2019;52:737-45.  Back to cited text no. 5
    
6.
Booth JR, Scheetz JP, Lemons JE, Eleazer PD. A comparison of torque required to fracture three different nickel-titanium rotary instruments around curves of the same angle but of different radius when bound at the tip. J Endod 2003;29:55-7.  Back to cited text no. 6
    
7.
Roland DD, Andelin WE, Browning DF, Hsu GH, Torabinejad M. The effect of preflaring on the rates of separation for 0.04 taper nickel titanium rotary instruments. J Endod 2002;28:543-5.  Back to cited text no. 7
    
8.
Alapati SB, Brantley WA, Svec TA, Powers JM, Nusstein JM, Daehn GS. Proposaed role of embedded dentin chips for the clinical failure of nickel-titanium rotary instruments. J Endod 2004;30:339-41.  Back to cited text no. 8
    
9.
Viana AC, Gonzalez BM, Buono VT, Bahia MG. Influence of sterilization on mechanical properties and fatigue resistance of nickel-titanium rotary endodontic instruments. Int Endod J 2006;39:709-15.  Back to cited text no. 9
    
10.
Schrader C, Peters OA. Analysis of torque and force with differently tapered rotary endodontic instruments in vitro. J Endod 2005;31:120-3.  Back to cited text no. 10
    
11.
Cheung GS, Darvell BW. Low-cycle fatigue of niTi rotary instruments of various cross-sectional shapes. Int Endod J 2007;40:626-32.  Back to cited text no. 11
    
12.
Gambarini G, Miccoli G, Seracchiani M, Khrenova T, Donfrancesco O, D'Angelo M, et al. Role of the flat-designed surface in improving the cyclic fatigue resistance of endodontic niTi rotary instruments. Materials (Basel) 2019;12: pii: E2523.  Back to cited text no. 12
    
13.
Bhagabati N, Yadav S, Talwar S. Anin vitro cyclic fatigue analysis of different endodontic nickel-titanium rotary instruments. J Endod 2012;38:515-8.  Back to cited text no. 13
    
14.
Saber SE. Factors influencing the fracture of rotary nickel titanium instruments. ENDO (Lond Engl) 2008;2:273-83.  Back to cited text no. 14
    
15.
Ferreira F, Adeodato C, Barbosa I, Aboud L, Scelza P, Zaccaro Scelza M, et al. Movement kinematics and cyclic fatigue of niTi rotary instruments: A systematic review. Int Endod J 2017;50:143-52.  Back to cited text no. 15
    
16.
De-Deus G, Moreira EJ, Lopes HP, Elias CN. Extended cyclic fatigue life of F2 proTaper instruments used in reciprocating movement. Int Endod J 2010;43:1063-8.  Back to cited text no. 16
    
17.
You SY, Bae KS, Baek SH, Kum KY, Shon WJ, Lee W. Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals. J Endod 2010;36:1991-4.  Back to cited text no. 17
    
18.
Gündoǧar M, Özyürek T. Cyclic fatigue resistance of oneShape, hyFlex EDM, WaveOne Gold, and reciproc blue nickel-titanium instruments. J Endod 2017;43:1192-6.  Back to cited text no. 18
    
19.
Özyürek T. Cyclic fatigue resistance of reciproc, waveOne, and WaveOne Gold nickel-titanium instruments. J Endod 2016;42:1536-9.  Back to cited text no. 19
    
20.
Mpro Brochure. Available from: http://www.imdmedical.com/#/Products/MPro%203%20Files%20System. [Last accessed on 2019 Oct 10].  Back to cited text no. 20
    
21.
Saber Sel D, Abu El Sadat SM. Effect of altering the reciprocation range on the fatigue life and the shaping ability of waveOne nickel-titanium instruments. J Endod 2013;39:685-8.  Back to cited text no. 21
    
22.
Ozyurek T, Uslu G, Yılmaz K. Influence of different movement kinematics on cyclic fatigue resistance of nickel-titanium instruments designed for retreatment. Saudi Endod J 2017;7:151-5.  Back to cited text no. 22
  [Full text]  
23.
Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod 2009;35:1469-76.  Back to cited text no. 23
    
24.
Hülsmann M, Donnermeyer D, Schäfer E. A critical appraisal of studies on cyclic fatigue resistance of engine-driven endodontic instruments. Int Endod J 2019;52:1427-45.  Back to cited text no. 24
    
25.
Jamleh A, Yahata Y, Ebihara A, Atmeh AR, Bakhsh T, Suda H, et al. Performance of niTi endodontic instrument under different temperatures. Odontology 2016;104:324-8.  Back to cited text no. 25
    
26.
de Vasconcelos RA, Murphy S, Carvalho CA, Govindjee RG, Govindjee S, Peters OA, et al. Evidence for reduced fatigue resistance of contemporary rotary instruments exposed to body temperature. J Endod 2016;42:782-7.  Back to cited text no. 26
    
27.
Dosanjh A, Paurazas S, Askar M. The effect of temperature on cyclic fatigue of nickel-titanium rotary endodontic instruments. J Endod 2017;43:823-6.  Back to cited text no. 27
    
28.
Cunningham WT, Balekjian AY. Effect of temperature on collagen-dissolving ability of sodium hypochlorite endodontic irrigant. Oral Surg Oral Med Oral Pathol 1980;49:175-7.  Back to cited text no. 28
    
29.
de Hemptinne F, Slaus G, Vandendael M, Jacquet W, De Moor RJ, Bottenberg P, et al. In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochlorite. J Endod 2015;41:1112-5.  Back to cited text no. 29
    
30.
Plotino G, Grande NM, Mercadé Bellido M, Testarelli L, Gambarini G. Influence of temperature on cyclic fatigue resistance of proTaper gold and proTaper universal rotary files. J Endod 2017;43:200-2.  Back to cited text no. 30
    
31.
Varela-Patiño P, Ibañez-Párraga A, Rivas-Mundiña B, Cantatore G, Otero XL, Martin-Biedma B, et al. Alternating versus continuous rotation: A comparative study of the effect on instrument life. J Endod 2010;36:157-9.  Back to cited text no. 31
    
32.
Karataş E, Arslan H, Büker M, Seçkin F, Çapar ID. Effect of movement kinematics on the cyclic fatigue resistance of nickel-titanium instruments. Int Endod J 2016;49:361-4.  Back to cited text no. 32
    
33.
Pirani C, Iacono F, Generali L, Sassatelli P, Nucci C, Lusvarghi L, et al. HyFlex EDM: Superficial features, metallurgical analysis and fatigue resistance of innovative electro discharge machined niTi rotary instruments. Int Endod J 2016;49:483-93.  Back to cited text no. 33
    
34.
Pedullà E, Lo Savio F, Boninelli S, Plotino G, Grande NM, La Rosa G, et al. Torsional and cyclic fatigue resistance of a new nickel-titanium instrument manufactured by electrical discharge machining. J Endod 2016;42:156-9.  Back to cited text no. 34
    
35.
Testarelli L, Plotino G, Al-Sudani D, Vincenzi V, Giansiracusa A, Grande NM, et al. Bending properties of a new nickel-titanium alloy with a lower percent by weight of nickel. J Endod 2011;37:1293-5.  Back to cited text no. 35
    
36.
Zhou HM, Shen Y, Zheng W, Li L, Zheng YF, Haapasalo M. Mechanical properties of controlled memory and superelastic nickel-titanium wires used in the manufacture of rotary endodontic instruments. J Endod 2012;38:1535-40.  Back to cited text no. 36
    
37.
Thompson SA. An overview of nickel-titanium alloys used in dentistry. Int Endod J 2000;33:297-310.  Back to cited text no. 37
    


    Figures

  [Figure 1], [Figure 2]
 
 
    Tables

  [Table 1]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Materials and Me...
Results
Discussion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed912    
    Printed26    
    Emailed0    
    PDF Downloaded98    
    Comments [Add]    

Recommend this journal